Superseded Standard
Historical

ASTM E2069-06

Standard Test Method for Temperature Calibration on Cooling of Differential Scanning Calorimeters

Summary

1.1 This test method covers the temperature calibration of differential scanning calorimeters on cooling using the difference between transition temperatures upon heating and cooling in the temperature range of 50 to 185°C. An offset in the indicated temperature between heating and cooling experiments, within this temperature range, may be used to provide temperature calibration on cooling at other temperature ranges.

1.2 SI units are the standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 6.


Significance and Use:

This test method is useful in calibrating the temperature signal of a differential scanning calorimeter for cooling experiments such as the determination of crystallization temperatures in Test Method D 3418 and Test Method E 794.

This test method may be used for research, development, analytical, specification acceptance, quality assurance and control purposes.

Technical characteristics

Publisher American Society for Testing and Materials (ASTM International)
Publication Date 08/15/2006
Collection
Page Count 7
Themes Calorimetry
EAN ---
ISBN ---
Weight (in grams) ---