Active Draft standard
Most Recent

ISO/DIS 10524-3:2025

Pressure regulators for use with medical gases
Part 3: Pressure regulators integrated with cylinder valves (VIPRs)

Summary

This document specifies design, type testing, and marking requirements for cylinder valves with integrated pressure regulators [as defined in 3.26 and referred to hereafter as valves with integrated pressure regulators (VIPRs)] intended for the administration of medical gases in the treatment, management, diagnostic evaluation and care of patients or for gases used for driving surgical tools. Examples of gases include oxygen, medical air and oxygen/nitrous oxide mixtures. This document applies to VIPRs mounted on refillable cylinders with a working pressure up to 30 000 kPa (300 bar) intended to be filled in cylinder filling facilities or on self-filling systems as used in homecare applications. VIPRs covered by this document are pressure pre-set and provided with a pressure outlet and/or pre-set flow outlet(s).

Notes

40.00 : DIS enregistré

Technical characteristics

Publisher International Organization for Standardization (ISO)
Publication Date 12/12/2025
Edition 3
Page Count 48
EAN ---
ISBN ---
Weight (in grams) ---