Withdrawn
Standard
Most Recent
NBN ISO 11171:2021
Hydraulic fluid power — Calibration of automatic particle counters for liquids
Summary
This document specifies procedures for the following:
a) primary particle-sizing calibration for particle sizes 1 µm(c) and larger, sensor resolution and
counting performance of liquid automatic particle counters that are capable of analysing bottle
samples;
b) secondary particle-sizing calibration using suspensions verified with a primary calibrated APC;
c) establishing acceptable operation and performance limits;
d) verifying particle sensor performance using a test dust;
e) determining coincidence and flow rate limits.
This document is applicable for use with hydraulic fluids, aviation and diesel fuels, engine oil and
other petroleum-based fluids. This document is not applicable to particle-sizing calibration using
NIST SRM 2806b primary calibration suspensions.
a) primary particle-sizing calibration for particle sizes 1 µm(c) and larger, sensor resolution and
counting performance of liquid automatic particle counters that are capable of analysing bottle
samples;
b) secondary particle-sizing calibration using suspensions verified with a primary calibrated APC;
c) establishing acceptable operation and performance limits;
d) verifying particle sensor performance using a test dust;
e) determining coincidence and flow rate limits.
This document is applicable for use with hydraulic fluids, aviation and diesel fuels, engine oil and
other petroleum-based fluids. This document is not applicable to particle-sizing calibration using
NIST SRM 2806b primary calibration suspensions.
Technical characteristics
| Publisher | Bureau de Normalisation Belge (NBN) |
| Publication Date | 06/17/2021 |
| Cancellation Date | 01/18/2023 |
| Page Count | 2 |
| EAN | --- |
| ISBN | --- |
| Weight (in grams) | --- |