Superseded Standard
Historical

NBN ISO 17664-2:2021

Processing of health care products — Information to be provided by the medical device manufacturer for the processing of medical devices — Part 2: Non-critical medical devices

Summary

This document specifies requirements for the information to be provided by the medical device
manufacturer for the processing of non-critical medical devices not intended to be sterilized (i.e.
a medical device that is intended to come into contact with intact skin only or a medical device not
intended for direct patient contact).
This includes information for processing prior to use or reuse of the medical device.
Processing instructions are not defined in this document. Rather, this document specifies requirements
to assist manufacturers of medical devices in providing detailed processing instructions that consist of
the following activities, where applicable:
a) preparation before processing;
b) cleaning;
c) disinfection;
d) drying;
e) inspection and maintenance;
f) packaging;
g) storage;
h) transportation.
This document excludes processing of:
1) critical and semi-critical medical devices;
2) medical devices intended to be sterilized;
3) textile medical devices used in patient draping systems or surgical clothing;
4) medical devices specified by the manufacturer for single use only and supplied ready for use.
NOTE See Annex E for further guidance on the application of the ISO 17664 series to a medical device.

Technical characteristics

Publisher Bureau de Normalisation Belge (NBN)
Publication Date 06/17/2021
Cancellation Date 12/21/2023
Page Count 32
EAN ---
ISBN ---
Weight (in grams) ---
No products.